贝克曼库尔特离心机 S7 系统故障的解决方案:目前,实验室内一台贝克曼库尔特离心机开机后不能启动,屏幕显示的错误代码为:“S7 System Error, Call for Service Please.”,考虑到这台离心机很少使用,有可能是主板上ROM内存在的信息丢失或错误。高速离心机BECKMANAvanti J-E上网查找一下解决方案,果然有重置ROM参数的方案如下:法一、 Enter 9 1 1 CE 操作面板上各键顺次按下即可。操作面板上如果没有 Enter 键,改用:法二、 Time Accel CE 键或法三、 Set Accel Set Decel Time Accel Clear 键如果您也遇到类似问题、祝您马到成功、工作顺利。原文参考如下:System Error S7View Beckman Coulter Avanti J Series Questions and Answers forum View Beckman Coulter Avanti J Series product details Q: At the start of a run theres a S7 system error that appears.A:The S7 Error is a system error where there may be some memory loss due to a power disruption in the building. Try pressing the Enter key then 911 then CE to clear the error and try another run. If problems persist after this then service will be required. I hope this helps. Michael Girard Product Support Services Beckman Coulter Inc. 1-800-369-0333 x353419(Thank You to Michael from Beckman Coulter for your response)Best Wishes,LabWrench Admin.Q: Im sorry, can you help me too? We have J-25 old centrifuge but it doesnt have ENTER key. How can I clear S7 error in this case?A:To clear an S7 on a J25; press Time, Accel, CE. All your calibration factors were set to Default (100) and the elevation is set to 0 meters by the S7. This is when RAM Memory doesnt agree with other parameters. Bad power can sometimes cause this. The latest version of Firmware does not reset the calibration factors to default. You may want to have the Beckman technician bring the latest firmware and calibrate your instrument. If you prefer, I can email you the procedure to calibrate the instrument yourself.Q: Thank you for your proposition, but I do not have the button CE, so I cant do this.I have only these buttons: Accel, Decel, Clear, RPM/RCF, Time/Hold, Start, Stop, Set/Actual.A:The CE (Clear Entry) and the Clear button are the same thing.Q: I have tried to do this, but nothing happened. I have also changed the BBRAMs in tower of centrifuge - nothing. Is it the last chance to replace electronic board in tower?A:On the J25 series, you must enter a pre-code sequence to enter the repair code. I assumed that you had this pre-code. Sorry. Here is the full sequence of keys that must be pressed to clear an S7. Set-Accel-Set-Decel-Time-Accel-Clear. Disregard the hyphens. I put those in as separators in this text. Q: I have tried to put this code Set-Accel-Set-Decel-Time-Accel-Clear and
温馨提示
- 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
- 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
- 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
- 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
- 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
- 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
- 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。
最新文档
- 宁夏工商职业技术学院《现代电气控制技术》2024-2025学年第一学期期末试卷
- 云南省鲁甸县第二中学2025年化学高一第一学期期中经典模拟试题含解析
- 江苏省南通市2025-2026学年高一上生物期末达标检测模拟试题含解析
- 科技创新对实体经济转型影响分析
- 聚乙烯双壁波纹管规格尺寸试验记录
- 防水涂料固体含量试验记录
- 公司技术部负责人安全生产目标责任书
- 专业建设工作总结报告
- 供应链成本管理研究以苏宁易购公司为例
- 《语言学》(汉语言文学本科专业核心课程研究导引教材)述评
- 期中测试卷(1-4单元试卷)(含答案)2025-2026上学期六年级数学上册(人教版)
- 起重工考试题库及答案
- 成人创伤性颅脑损伤院前急救中国专家共识解读 3
- 武汉大学保密管理办法
- 铁路货运安全培训课件
- 混凝土公司年会活动方案
- 小学生世界艾滋病日宣传教育
- 2025年河北大学版(2024)小学信息科技三年级(全一册)教学设计(附目录 P179)
- 5Why原因分析方法培训
- 2025年4月保险学考试题+答案(附解析)
- 【安全经验分享】100例事故案例
评论
0/150
提交评论