1、If You've Exposed Your Paradigm Pump to an MRI, You Need to Read ThisLinda von WartburgJul 20, 2007 If your Paradigm pump has been not been exposed to powerful magnetic fields, such as those found near MRI machines, you have nothing to worry about. Go about your merry way and keep up the good wo

2、rk. In the unlikely event that such exposure has occurred, however, you need to be aware that it may cause the pump's motor to malfunction and significantly over-deliver insulin, causing severe hypoglycemia. As a result of this potential, the FDA has, based upon Medtronic's voluntary report,

3、 identified certain Paradigm pumps as subject to a Class II recall. (The action is described by Medtronic as a Class II field correction rather than a recall because no pump need be returned unless it has been actually damaged by magnetic exposure.) Medtronic reports that on April 24, 2007, the comp

4、any sent a warning letter, enclosed in a red envelope, to 200,000 Paradigm pump purchasers, as well to 30,000 healthcare practitioners, including every prescriber of their pumps, and every one of the over 14,000 MRI facilities in the United States. The patient letter read as follows: Recently, we ha

5、ve become aware of cases where exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls the movement of the motor used in Medtronic MiniMed Paradigm insulin infusion pumps. Although the insulin pumps exhibited alarms as a result of this damage,

6、in some cases, users have cleared these alarms and continued to use the pump. Under these conditions, the insulin pump will exhibit significant over-delivery of insulin, potentially causing severe hypoglycemia. Medtronic MiniMed (along with other insulin infusion pump manufacturers) has always cauti

7、oned users to avoid exposing their infusion pump to strong magnetic fields, such as those associated with MRI machines. Until recently, all testing has indicated this exposure would cause significant pump damage and prevent MiniMed Paradigm insulin pumps from delivering any insulin following this ex

8、posure. However, we now know that significant over-delivery of insulin can occur following this exposure. In light of this new information, we are reminding insulin pump users not to expose your insulin pump to strong magnetic fields such as those generated by MRI equipment. YOUR INSULIN PUMP MUST B

9、E REMOVED AND KEPT OUTSIDE THE PROCEDURE ROOM IF YOU ARE UNDERGOING MRI. If the insulin pump is damaged by exposure to a high strength magnetic field, the insulin pump will typically provide multiple and frequent alarms such as "motor error" alarms (especially after rewind), A35 alarms and

10、 A43 alarms. In addition, the pump will no longer provide a "low reservoir" alert prior to an empty reservoir and the actual volume in the reservoir will be less than the amount displayed on the pump's status screen. IF YOU SUSPECT THE PUMP MAY HAVE BEEN EXPOSED TO MRI OR OTHER VERY ST

11、RONG MAGNETIC FIELD, DISCONTINUE USE OF THE PUMP AND CONTACT THE MEDTRONIC MINIMED 24-HOUR HELP LINE AT 800-826-2099 FOR FURTHER ASSISTANCE. Please note that MiniMed Paradigm insulin infusion pumps have undergone extensive testing with devices such as airport metal detectors, electronic article surv

12、eillance equipment and cellular phones and we have confirmed that exposure to these devices will not adversely affect pump operation. If you have any additional questions regarding this letter or any other questions regarding our products, please feel free to contact the Medtronic MiniMed 24-Hour He

13、lp Line at 800-826-2099. Nearly three months later, in its July 11, 2007, weekly Enforcement Report, the FDA announced that Medtronic had initiated action regarding Paradigm pumps that the FDA categorized as a Class II recall. The FDA report read as follows: PRODUCT: Medtronic MiniMed Paradigm Insul

14、in Infusion Pumps for the following Paradigm models: MMT-511, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722, Recall # Z-1007-2007 CODE: All serial numbers for the following Paradigm models: MMT-51 1, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722 RECALLING FIRM/MANUFACTURER: Medtr

15、onic MiniMed, Northridge, CA, by letters on April 24, 2007. Firm initiated recall is ongoing. REASON: Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there we

16、re alarms as a result of the damage, some users cleared these alarms and continued using the pump. Under such conditions, the pump will significantly over-deliver; potentially causing severe hypoglycemia. VOLUME OF PRODUCT IN COMMERCE: 334,000 pumps DISTRIBUTION: Nationwide and Internationally On Ju

17、ly 14, Medtronic posted an announcement on the website Insulin-P, stating the following: "Recently, Medtronic became aware of cases where exposure to strong magnetic fields like Magnetic Resonance Imaging (MRI) resulted in damage to Medtronic Paradigm insulin infusion pumps. Medtronic

18、 has taken this situation very seriously, and we have voluntarily notified our patients, physicians and the FDA about these incidences, reinforcing the need for caution when using our insulin infusion pumps while in close proximity to strong magnetic fields like MRIs. The FDA has agreed with our ass

19、essment of the situation, and categorized this announcement as a Class II Field Correction, requiring us only to notify patients of the potential risk MRIs pose to their insulin pumps. Patients should not be concerned or take any further action unless they believe their Paradigm insulin pump has bee

20、n exposed to an MRI machine. Patients should not return their Paradigm insulin pump to Medtronic unless they believe this exposure has taken place. Extensive testing with other magnetized devices such as airport metal detectors, electronic article surveillance equipment and cellular phones will not

21、adversely affect pump operation." Medtronic asserts that it has handled the situation more than aggressively by intensifying its already strong warnings against exposing pumps to magnetic fields. We did have a few concerns, however. For one, why is there no fail-safe mechanism that automaticall

22、y shuts down the pump in the event of magnetic damage that might cause insulin over-delivery? As the Medtronic letter states, the pump does sound an alarm, but it is a generic alarm that pump users may be inclined to simply ignore. To address our concerns, we spoke with Medtronic and contacted the F

23、DA. Steve Sabicer of Medtronic notes that since May, when Medtronic discovered the potential for over-delivery of insulin, the company has not only voluntarily notified all its customers and distributors (as well as the FDA), but has also included a symbol on each pump that warns of exposure to magn

24、etic fields. With regard to the absence of a fail-safe feature, Mr. Sabicer states that Medtronic will work to design future insulin pump models to ensure they cannot over-deliver insulin in the event of magnetic exposure. In the meantime, he emphasizes, exposure to strong magnetic fields is forbidd

25、en for insulin pumps, and patients should contact Medtronic if they suspect such exposure. We were also concerned by the fact that according to the FDA's records, five adverse events related to the MRI issue were recorded after Medtronic's letters were sent. In that regard, Mr. Sabicer repor

26、ts that three of the five pumps in question passed all in-house inspections and were returned to their owners. In the other two cases, both patients exhibited hypoglycemia with no long-term adverse effects. Although the hypoglycemic events could not be definitively attributed to MRI exposure or even

27、 to the insulin pumps, the pumps were repaired or replaced. Based upon information from the FDA, the word "recall" is rather misleading in this case because it seems to imply that all pumps must be returned. In truth, however, the word "recall" is used by the FDA to announce a va

28、riety of far less drastic actions. As the FDA website states, "A medical device recall does not always mean that you must stop using the product or return it to the company. A recall sometimes means that the medical device needs to be checked, adjusted, or fixed.A recall is either a correction or a removal depending on where the action takes place. A correction addresses a problem with a medical device in the place where it is used or sold." A

温馨提示

  • 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
  • 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
  • 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
  • 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
  • 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
  • 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
  • 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。

评论

0/150

提交评论