下载本文档
版权说明:本文档由用户提供并上传,收益归属内容提供方,若内容存在侵权,请进行举报或认领
文档简介
1、S-600系列水汽取样装置S-600Series Water and Steam Sampling Device操作指南Op er at i on Man ua I版 号:01Edition No:01文件编号:ZG/CZ002-2006Document Number:ZG/CZ002-2006中美合资苏州赛华仪控有限公司SINO-US SUZHOU SAIHUA INSTRUMENT CONTROL CO.,LTD二oo六年二月02.200611 范围2.1.1Range本手册给出了 S-600 水汽取样装置的操作说明This manual introduces the operation
2、of the S-600 water and steam device.本手册适用于 S-600 水汽取样装置在现场的启动、停运和日常操作This manual is suitable for field startup, stoppage and routine work of theS-600 water and steam device.2.1.2装置启动前操作2 The operation before startup2.1降温架3.2.1Cooling frame5.1.1所有高压阀门应处于关闭状态。5.2.1 All the high pressure valves need to
3、 be close.5.2.2所有各连接接头、紧固件不应有松动、脱落现象。5.1.2 All conjunctions and fasteners shouldn t be loosen and be falloff.5.1.3恒压装置压力调节螺母旋至中位,出口调节阀处于关断状态。5.2.3 Put the pressure adjusting nut of constant pressure device at median, andclose the outer adjusting valves.5.2.4打开冷却水进出口母管总阀,冷却水进口母管的压力应大于0.2MPa。25.1.4 Op
4、en the master valve of cooling water in-and-out parent pipes and thepressure should over than 0.2MPa。3.2.2 人工取样盘、仪表盘3.3.2Artificial sampling boards, dash boards3.3.2检查所有阀门及部件连接处是否连接可靠。5.3.1 Check all the conjunctions between valves and parts firmly5.3.2人工取样盘上的限流阀应处于最大开度。3.3.2 The current-limiting va
5、lves of artificial sampling boards need to be largestopen.3.3.2关闭仪表盘上所有仪表取样阀门。5.3.3 Closed all the meter-sampling valves of dash boards.3.3.2 装置闭环保护系统3.2.3Closed device of protecting system5.3.1温控开关Temperature controlling switches 温控开关通电后输出闭合信号,保护装置应实施保护动作。 When the temperature controlling switches
6、power on and outputthe closing signal, the protecting device run.靶式流量控制器5.3.2 Target style flow-controller.关闭冷却水进水阀门或出水阀门, 靶式流量控制器应输出闭合信号, 各样水保护装置应实施保护动作。同时3Close the cooling water in-and-out valves or outer valves, target styleflow-controller outputs the closing signal, and protectingdevices run.5.
7、5.2化学分析仪表3.2.4Chemical analyze meter 仪表盘电气箱内电器元件齐全, 接线端子连接可靠,分别接通各仪表电 源,仪表接线正确。The electric appliances of dash boards component inside the electricity box iswell-found, connecting line to carry subconjunction credibility, connecting eachappearance power supply respectively, the meter connects line e
8、xactitude.3 装置启动2.1.3.Device startup2.1.3降温架Cooling frame依次开启冷却器的进出口球阀。Open the in-and out ball valves of cooling devices one by one in order.返冲过滤器置于“运行”状态。The recoiling filter is on running.逐路开启装置样水进口一次门和二次门(全开) ,打开排污门到最大开 度,对各路样水逐路进行排污。Openthe first inner door of sampling water device, and then th
9、e second one(full), and open the blow-off door largest for draining.4排污结束,关闭排污门。Turn off the blow-off doors after draining.打开恒压装置出口关断阀Open stopped valves of constant pressure devices.人工取样盘2.1.3Artificial sampling boards2.2.3开启人工取样盘样水门。Open the door of artificial sampling boards.调出人工取样流量。2.2.3 Analyz
10、e the flow rate of artificial sampling.2.1.3仪表盘Dash boards取出各仪表样水过滤器滤芯。Get out filter elements of each sampling water meter.从各发送器内取出电极。Take out the poles from each transmitter.开启仪表盘上各入口节流阀,对仪表样水管路进行冲洗,时间 1 分钟。 冲洗结束后,重新安装好样水过滤器滤芯和发送器电极。Openthe inlet throttles of dash boards and wash pipes by 1 minute
11、, then fullinstall them.全开恒压装置出口调节阀根据每路样水所配仪表的数量,调节恒压装 置调节5螺母,使样水总流量符合要求。According to the quantity of sampling water meter, full open the outeradjusting valves of constant pressure devices, while as to adjust the nuts formeeting the request.注:人工调节流量为500ml/min , DD PH 表均为 300ml/min , O2表500ml/min , 其
12、它表计 150ml/min 。PS: the flow rate of manpoweris 500ml/min , DD、PHare 300ml/min , O2meter is 500ml/min, other meters are 150ml/min 。洀戀攀爀攀搀开戀挀攀 愀攀 愀 昀昀愀愀攀昀 一甀洀戀攀爀攀搀开挀昀 攀攀愀 愀 愀昀昀 愀搀戀攀搀挀 一甀洀戀攀爀攀搀开挀戀 昀戀 搀 挀攀愀挀搀昀攀 一甀洀戀攀爀攀搀开戀昀搀据流量计的指示值,调节流量计入口节流阀,使各仪表样水流量符 合 3.3.4 中各仪表的要求。Adjust the inner throttle of flow m
13、eter according to the designation of theflow meter.恒温装置2.1.3Constant temperature devices恒温装置启动操作详见SW 療列恒温装置使用手册。The operation of constant devices details were introduced in the operationmanual of SWJ series constant devices.2.1.4装置停运4 .Stoppage devices4.1 关闭降温架样水进口一次门、二次门。6Close the sampling water i
14、nner first door and second door of cooling frame.关闭仪表盘节流阀。4.2 Close the throttle of dash boards.4.3 关闭温控电源、仪表电源、恒温装置电源。Close the power of temperature controlling, meter and constant devices.关闭装置的冷却水源。Close the cooling headwater of devices.5 装置运行注意事项5. Matters needing attention5.1 高压阀high pressure va
15、lves高压阀门必须处于全开状态或全关状态,禁止用作节流。The high pressure valves should to be full open or full closed, and which isforbidden for cutting flow.打开或关闭高压阀时,必须使用专用工具扭力扳手,按高压阀操作指 南要求的扭力矩进行操作。5.1.2It should use special tension wrench when open and close the high pressurevalves and according to the operation manual o
16、f high pressure valves.筒形冷却器5.2 Tube-shaped cooler5.2.1冷却器冷却水进出口球阀必须全开。7The inner and outer ball valves of cooling water of cooling devices should be open.5.3 高压过滤器high pressure filters正常取样Sampling正常运行时高压过滤器手轮处于水平“运行”状态The high pressure filters are running when it on error-free running period.返冲清洗Re
17、coiling wash返冲清洗规定时间为每 15 天一次。返冲清洗前,先打开排污阀, 1 分钟 后,逆时针 90旋转高压过滤器手轮,使手轮处于垂直“返冲”状态,返冲 清洗时间 15秒。冲洗结束后, 旋转手轮处于水平“运行” 状态,关闭排污阀。Recoil washing should be done in each 15 days. Open the blow-off valvesbefore washing, then get the hand wheel of high pressure filter counterclockwiserotation 90after 1 minute,
18、to make sure it is onrecoil situation 15 seconds. Then the rotary hand wheel should be on running andclose the blow-off valves after finishing. 注:在机组启动初期,不准进行返冲清洗。PS: It can not to recoiling wash at the beginning of start-up高压排污5.4 high-pressure draining 冲管排污规定时间为每周一次。排污时,先关闭恒压装置出口调节阀, 再打开高压排污阀,进行正常
19、排污,时间 1 分钟,排污结束后,关8闭高压排 污门,打开恒压装置出口调节阀。Washing pipes drain in each one week. Close the outer adjusting valves ofconstant pressure devices and then open high pressure blow-off valves beforedraining, and hold on 1 minute. Close highpressure blow-off door and open the outer adjusting valves of constant
20、 pressuredevices when it is finished.注:必须逐路排污,严禁同时多路排污。PS: It must be drained one by one.5.5 压力、流量的调节adjusting of pressure and flow压力、流量调节Adjusting of pressure and flow缓慢调节恒压装置调节螺母,使压力在0.25Mpa 至 0.38Mpa 之间。根据每路样水所配仪表的数量,使样水总流量符合要求。在压力、流量均允许范围内 调定。Adjust the adjusting nuts of constant devices slowly,
21、 and be sure the pressure isbetween 0.25Mp to 0.38Mpa. To makesure the total sampling water flow rate ismeet the requirement according to the quantity of sampling water meter, and adjustthem under the suitable range of pressure and flow rate.注:人工取样 500ml/min ; DD PH 表均为 300ml/min ; O2表 500ml/min ; 其
22、 它表计 150ml/min 。总流量为上述流量和的 1 至 1.2 倍。PS: the flow rate of manpower is 500ml/min; DD, PH are 300ml/min; O29meter is 500ml/min; other meters are 150ml/min. The total discharge is anabove-mentioned discharge with of 100%-120%.流量偏小时采取的措施The measures of the flow small than normal对高压过滤器进行返冲排污,清除滤芯过滤孔中杂质,增
23、加通流量。Drain the high filter and clean the impurity of filter element for adding through-flowrate.缓慢调节恒压装置调节螺母。Adjust the adjusting nuts of constant devices slowly.日常调节Normal adjusting按 4.1 要求每周调节一次。According to 4.1 says once a week.保护功能检查5.6 check the protecting system 恒压装置具有自锁保护功能,即恒压装置后端保护装置因某种原因动作 保护后,恒压装置因样水系统压力升高而自动锁闭样水通道,故每星期要定 期进行一次冷却水断水或样水超温保护试验,以便检查保护装置的可靠性及 恒压装置的自锁功能。The constant pressure device had protecting function of self-locking, whichnamely constant pressure de
温馨提示
- 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
- 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
- 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
- 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
- 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
- 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
- 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。
评论
0/150
提交评论