版权说明:本文档由用户提供并上传,收益归属内容提供方,若内容存在侵权,请进行举报或认领
文档简介
GB/T30099—2025
目次
前言··························································································································Ⅲ
1范围·······················································································································1
2规范性引用文件········································································································1
3术语和定义··············································································································1
4产品分类·················································································································2
5工作条件·················································································································2
6要求·······················································································································3
7试验方法·················································································································7
8检验规则···············································································································11
9标识·····················································································································13
10包装、运输及贮存···································································································13
Ⅰ
GB/T30099—2025
前言
本文件按照GB/T1.1—2020《标准化工作导则第1部分:标准化文件的结构和起草规则》的规
定起草。
本文件代替GB/T30099—2013《实验室离心机通用技术条件》,与GB/T30099—2013相比,除编
辑性改动外,主要技术变化如下:
——更改了标准的中文名称和英文名称,将中文名称“实验室离心机通用技术条件”改为“实验室
离心机”,英文名称“Generalspecificationsforlaboratorycentrifuge”改为“Laboratorycentri⁃
fuges”;
——增加了术语条目“冷冻离心机”“制热离心机”和“常温离心机”,删除了术语条目“超速离心机”
“低速冷冻离心机”和“高速冷冻离心机”,更改了术语条目“立式离心机”(见第3章,2013年
版的第3章);
——更改了产品分类(见第4章,2013年版第4章);
——更改了“5使用条件”为“5工作条件”(见第5章,2013年版第5章);
——增加了对结构、振幅、制热、机械安全和电磁兼容性的要求(分别见6.2、6.8、6.10、6.11.2
和6.12);
——更改了对转速相对误差、转速稳定性、试液温升、制冷温度误差和联锁保护的要求(分别见
6.3、6.4、6.6、6.9.3和6.11.2.5,2013年版的6.3、6.4、6.6、6.8.3和6.9.5);
——更改了试验条件(见7.2,2013年版的7.2);
——增加了结构、振幅、制热、机械安全和电磁兼容性的试验方法(分别见7.4、7.10、7.12、7.13.2
和7.14);
——删除了旋转组件重量偏差和噪声声功率级的试验方法(分别见2013年版的7.4和7.7.2);
——更改了噪声发射声压级、试液温升和制冷的试验方法(分别见7.7、7.8和7.11,2013年版的
7.7.1、7.8和7.10);
——删除了“周期检验”(见2013年版的8.5);
——增加了对警示标志的要求(见9.2)。
请注意本文件的某些内容可能涉及专利。本文件的发布机构不承担识别专利的责任。
本文件由中国机械工业联合会提出。
本文件由全国实验室仪器和设备标准化技术委员会(SAC/TC526)归口。
本文件起草单位:湖南省计量检测研究院、湖南赫西仪器装备有限公司、湖南湘仪实验室仪器开发
有限公司、长沙英泰仪器有限公司、上海安亭科学仪器厂、机械工业仪器仪表综合技术经济研究所、
长沙湘智离心机仪器有限公司、长沙迈佳森仪器设备有限公司、湖南德诺科学仪器有限公司、上海天美
生化仪器设备工程有限公司、岳阳市检验检测中心、湖南平科科学仪器有限公司、湖南湘鑫仪器仪表有
限公司、湖南润坤检测技术有限公司、北京白洋医疗器械有限公司、湖南声仪测控科技有限责任公司、
青岛海尔生物医疗股份有限公司、新疆维吾尔自治区计量测试研究院、江苏拓米洛高端装备股份有限
公司。
本文件主要起草人:陈庆雄、王凯、魏强、徐昌平、张桂玲、寻继勇、田健、周雄、朱宪宇、罗建明、
孙华、练达、刘凯、李肈星、沈利良、蔡荤、卞挚诚、姚希华、李晓南、肖风、张小芹、武育荣、熊知明、刘志杰、
刘湘衡、李正生、郭丽、张艳军。
Ⅲ
GB/T30099—2025
本文件及其所替代文件的历次版本发布情况为:
——GB/T30099—2013;
——本次为第一次修订。
Ⅳ
GB/T30099—2025
实验室离心机
1范围
本文件界定了实验室离心机的术语和定义,规定了实验室离心机的产品分类、工作条件、要求、试
验方法、检验规则、标识、包装、运输及贮存。
本文件适用于最高额定转速不超过30000r/min的实验室离心机(以下简称离心机)的设计、生产
及检验。
2规范性引用文件
下列文件中的内容通过文中的规范性引用而构成本文件必不可少的条款。其中,注日期的引用文
件,仅该日期对应的版本适用于本文件;不注日期的引用文件,其最新版本(包括所有的修改单)适用于
本文件。
GB/T191包装储运图示标志
GB4793.1—2007测量、控制和实验室用电气设备的安全要求第1部分;通用要求
GB4824工业、科学和医疗设备射频骚扰特性限值和测量方法
GB/T14710医用电器环境要求及试验方法
GB/T17248.2—2018声学机器和设备发射的噪声在一个反射面上方可忽略环境修正的近
似自由场测定工作位置和其他指定位置的发射声压级
GB17625.1电磁兼容限值第1部分:谐波电流发射限值(设备每相输入电流≤16A)
GB/T17625.2电磁兼容限值对每相额定电流≤16A且无条件接入的设备在公用低压供电
系统中产生的电压变化、电压波动和闪烁的限制
GB/T17626.2电磁兼容试验和测量技术静电放电抗扰度试验
GB/T17626.3电磁兼容试验和测量技术第3部分:射频电磁场辐射抗扰度试验
GB/T17626.4电磁兼容试验和测量技术电快速瞬变脉冲群抗扰度试验
GB/T17626.5电磁兼容试验和测量技术浪涌(冲击)抗扰度试验
GB/T17626.6电磁兼容试验和测量技术射频场感应的传导骚扰抗扰度
GB/T17626.8电磁兼容试验和测量技术工频磁场抗扰度试验
GB/T17626.11电磁兼容试验和测量技术第11部分:对每相输入电流小于或等于16A设
备的电压暂降、短时中断和电压变化抗扰度试验
3术语和定义
下列术语和定义适用于本文件。
3.1
实验室离心机laboratorycentrifuge
能对样品材料施加离心作用的实验室用仪器。
注:适用于科研、制药、医用等领域的实验室。
1
GB/T30099—2025
3.2
台式离心机bench⁃topcentrifuge
置于实验台(桌)面操作使用的离心机。
3.3
落地式离心机floortypecentrifuge
置于地面操作使用的离心机。
3.4
低速离心机lowspeedcentrifuge
最高额定转速小于10000r/min的离心机。
3.5
高速离心机highspeedcentrifuge
最高额定转速不小于10000r/min且不超过30000r/min的离心机。
3.6
冷冻离心机refrigeratedcentrifuge
带有制冷系统的离心机。
3.7
制热离心机heatingcentrifuge
仅带制热系统的离心机。
3.8
常温离心机normaltemperaturecentrifuge
不带制冷或制热系统的离心机。
4产品分类
4.1离心机按结构分为:台式离心机和落地式离心机。
4.2离心机按转速分为:低速离心机和高速离心机。
4.3离心机按功能分为:冷冻离心机、制热离心机和常温离心机。
5工作条件
5.1环境条件
离心机正常工作环境如下:
a)温度:5℃~35℃;
b)相对湿度:≤80%;
c)大气压强:86kPa~106kPa;
d)无阳光直接照射或其他热源直接辐射;
e)污染等级不超过2级,周围无导电尘埃、爆炸性气体、高浓度粉尘或腐蚀性物质存在;
f)周围无强烈振动和气流存在。
5.2供电条件
离心机供电电源应符合以下条件:
a)额定电压交流220V或380V,允许偏离额定值的范围为±10%;
2
GB/T30099—2025
b)额定频率50Hz,允许偏离额定值的范围为±2%。
6要求
6.1外观
6.1.1离心机表面应平整,文字和标识应清晰。
6.1.2离心机的涂镀层应色泽均匀,无露底、脱皮、起泡、毛刺、斑痕及明显的划痕。
6.2结构
6.2.1机架
机架应具有足够的强度,满足运行安全需求。
6.2.2机盖
6.2.2.1机盖上应设置透明观察窗,透明观察窗宜设置在离心腔的上方。
6.2.2.2机盖应具备足够的机械强度,起到安全防护的作用。
6.2.3旋转组件
6.2.3.1转头及其附件应具有足够的强度,满足存放样品及自身离心力的要求。
6.2.3.2与离心腔壁应保持足够的安全距离。
6.2.4有特殊要求样品的密封
6.2.4.1机盖与离心腔、离心腔与驱动系统之间均应设置密封装置。
6.2.4.2旋转组件宜设置生物密封装置,并满足高温灭菌的要求。
6.3转速相对误差
工作条件下,离心机所配置的任意一个转头以最大载荷和最高额定转速运行时,测得的转速与最
高额定转速的相对误差,对于低速离心机不超过±1.0%,高速离心机不超过±0.3%。
6.4转速稳定性
工作条件下,离心机所配置的任意一个转头以最大载荷和最高额定转速运行时,转速的不稳定性,
对于低速离心机不超过±1.0%,高速离心机不超过±0.3%。
6.5噪声
工作条件下,离心机所配置的任意一个转头以最大载荷和最高额定转速运行时的噪声,在操作者
(或观察者)位置和距离离心机表面1m处的A计权发射声压级不应超过70dB。
如果超过该限值,制造商应在其技术文件中要求使用设备的责任部门采取防护措施。
6.6试液温升
工作条件下,常温离心机所配置的任意一个转头以最大载荷和最高额定转速运行,在运转表1规
定的时间后,离心管(瓶)内试液温升应符合表1的要求。
3
GB/T30099—2025
表1
类型运转时间/min试液温升/℃
高速离心机15≤12
低速离心机20≤12
6.7升速时间和降速时间
离心机的升速时间和降速时间应符合表2的要求。
表2
升速时间降速时间
类别
总容量<3000mL总容量≥3000mL总容量<3000mL总容量≥3000mL
低速离心机≤3min≤5min≤5min≤7min
高速离心机≤4min≤5min≤6min≤7min
6.8振幅
工作条件下,离心机所配置的任意一个转头以最大载荷和最高额定转速运行时,运行应平稳,振幅
应不超过0.1mm。
6.9制冷
6.9.1通则
6.9中的各条只适用于冷冻离心机(不包括带真空系统的冷冻离心机)。
6.9.2制冷时间
环境温度在(25±2)℃的条件下,离心机不预冷,当离心机所配置的任意一个转头以最大载荷和最
高额定转速运行时,离心腔内温度从25℃降到4℃的制冷时间应不超过20min。
6.9.3温度误差
环境温度在(25±2)℃的条件下,离心腔温度控制范围在0℃~25℃时,温度误差不应超过±2℃。
6.10制热
6.10.1通则
6.10中的各条只适用于制热离心机。
6.10.2制热时间
环境温度在(25±2)℃的条件下,离心机不预热,当离心机所配置的任意一个转头以最大载荷和最
高额定转速运行时,离心腔内温度从25℃升到80℃的制热时间不应超过30min。
4
GB/T30099—2025
6.10.3温度误差
环境温度在(25±2)℃的条件下,离心腔温度控制范围在60℃~80℃时,温度误差不应超过±5℃。
6.11安全
6.11.1电气安全
6.11.1.1泄漏电流
离心机起防电击作用的电气隔离应有良好的性能,以使流过它的电流被限制在下列范围内:
a)外壳漏电流的容许值在正常状态下为100μA,单一故障状态下为500μA;
b)对地漏电流的容许值在正常状态下为0.5mA,单一故障状态下为3.5mA。
6.11.1.2介电强度
离心机应耐受以下介电强度试验:
a)电源L/N极和金属箔覆盖的易触及非金属操作面板之间,试验交流电压4000V,持续时间
1min,不能被击穿;
b)电源L/N极和已保护接地的可触及金属表面之间,试验交流电压1500V,持续时间1min,
不能被击穿。
6.11.1.3插头连接设备的保护连接阻抗
对于采用插头连接电源的离心机,保护导体端子与规定要采用保护连接的每一个可触及零部件之
间的阻抗不应超过0.1Ω,电源线的阻抗不构成规定的保护连接阻抗的一部分。
6.11.2机械安全
6.11.2.1正常运行位移
在正常使用过程中,离心机不应有相对于其安装位置目视可见的位移。
6.11.2.2机体抗倾倒
6.11.2.2.1离心机向任意方向倾斜10°,不应倾倒。
6.11.2.2.2对高度大于或等于1m且质量大于或等于25kg的离心机,以及所有落地式离心机,要在
其顶部施加一个力,该力为250N或离心机重量的20%,取其较小者。力沿除向上的所有方向施加。
离心机不应倾倒。
6.11.2.3非正常运行位移
离心机按照制造商的规定安装后,由转头组件不平衡造成的离心机移动、转头组件破裂、驱动系统
失效或锁死状况下,门盖不能打开,且任何尺寸超过1.5mm的旋转组件的零部件或碎片不应完全穿透
保护罩,较小尺寸的(烟雾和液体除外)应控制在安全空间内,即任何方向距离离心机最外端的零部件
300mm范围内。
6.11.2.4运动部件防护
离心腔顶部不应有直径4mm销钉可穿过的任何开孔。
机盖在转头驱动装置通电时应保持锁紧,直到转头组件转速为0时为止。在电源失效时,机盖锁
5
GB/T30099—2025
紧机构不应释放,应使用特殊工具才能打开。
在正常条件下,转头组件或驱动系统运动零部件应提供适当的防护措施以防止机械伤害。
6.11.2.5联锁保护
离心机应有机盖自锁功能。机盖打开时,离心机不能运转;合上机盖,离心机运转时,机盖不能被
打开;遇停电或故障状态时,应有开锁装置并只有使用工具才能打开机盖。
6.11.2.6不平衡保护
高速离心机、容量大于3000mL的低速离心机应具有不平衡保护功能,当达到一定不平衡量时,
应自动停机。
注:不平衡量由生产商自行规定。
6.12电磁兼容性
6.12.1发射要求
GB4824的规定适用于离心机的传导发射和辐射发射要求,设备的分类和限值的选择应在考虑预
期使用环境和使用区域的发射要求后确定。
GB17625.1的规定适用于离心机的谐波电流发射要求,设备的分类为标准中定义的A类设备。
GB/T17625.2的规定适用于离心机的电压变化、电压波动和闪烁的限制要求,设备的分类和限值
的选择应在考虑预期使用环境和使用区域的发射要求后确定。
6.12.2抗扰度要求
离心机的抗扰度要求在表3中给出。
表3
序号试验项目试验要求基础标准性能评定
1静电放电抗扰度接触放电4kV、空气放电8kVGB/T17626.2B
3V/m(80MHz~1GHz)
2射频电磁场辐射抗扰度3V/m(1.4GHz~2GHz)GB/T17626.3A
1V/m(2.0GHz~2.7GHz)
3电快速瞬变脉冲群抗扰度电源端口:1kV(5ns/50ns,5kHz)GB/T17626.4B
4浪涌(冲击)抗扰度电源端口:线对线1kV,线对地2kVGB/T17626.5B
5射频感应的传导骚扰抗扰度3V(150kHz~80MHz)GB/T17626.6A
6工频磁场抗扰度试验30A/mGB/T17626.8A
剩余电压<5%,持续0.5个周期B
电压暂降
7剩余电压<70%,持续25个周期GB/T17626.11B
短时中断剩余电压<5%,持续250个周期C
6.13环境适应性
离心机的环境适应性要求在表4和表5中给出。
6
GB/T30099—2025
表4
试验要求
序号试验项目持续恢复性能评定
试验条件通电状态
时间时间
1额定工作低温试验5℃≥1h—试验时通电试验中应符合6.3要求
2低温贮存试验-40℃4h4h试验后通电试验后应符合6.3要求
3额定工作高温试验40℃≥1h—试验时通电试验中应符合6.3要求
4高温贮存试验55℃4h4h试验后通电试验后应符合6.3要求
30℃,相对湿度
5额定工作湿热试验≥4h8h试验后通电试验后应符合6.3要求
(93±3)%
40℃,相对湿度
6湿热贮存试验48h24h试验后通电试验后应符合6.3要求
(93±3)%
表5
试验要求
序号试验项目性能评定
试验条件通电状态
频率循环范围(5—20—5)Hz;
振幅值0.15mm;
1振动试验试验后通电试验后应符合6.3要求
扫频次数为10次;
扫频速率≤1倍频程每分
加正弦波、频率
加速度为50m/s2;
2碰撞试验试验后通电试验后应符合6.3要求
脉冲持续时间(11±2)ms;
碰撞次数为(1000±10)次
7试验方法
7.1主要试验仪器及设备
主要试验仪器及设备的名称和要求如下:
a)天平,检定分度值不高于0.1g;
b)转速表,准确度等级不低于0.5级;
c)测温仪,最大允许误差为±0.5℃;
d)电子秒表,最大允许误差为±1s/d;
e)声级计,性能等级不低于2级;
f)测振仪,幅值线性度最大允许误差为±10%;
g)非常规检验时所用的有关仪器设备。
7.2试验环境和供电条件
离心机的试验环境和供电条件应满足下列要求:
7
GB/T30099—2025
a)环境温度:15℃~30℃;
b)相对湿度:≤80%;
c)大气压强:86kPa~106kPa;
d)交流供电电源:额定电压为220V或380V,允许偏离额定值的范围为±10%;额定频率为50Hz,
允许偏离额定值的范围为±2%;总谐波失真小于5%。
7.3外观
目视检查。
7.4结构
目视检查。
7.5转速相对误差
将装至满载转头的离心机调至其转头对应的最高额定转速,达到最高额定转速运行5min。用转
速表测量离心机的转速,每隔1min测1次,共测量5次,按公式(1)计算每一次的转速相对误差。
n-n
=imax×100%…………………(1)
δni
nmax
式中:
第次转速相对偏差;
δni——i
第次转速的测得值,单位为转每分(r/min);
ni——i
r/min
nmax——最高额定转速,单位为转每分()。
7.6转速稳定性
取7.5的测试数据,计算5次转速的算术平均值,并按公式(2)计算每一次的转速不稳定性,取最
大值。
|n-nˉ|
δ=i×100%…………………(2)
sinˉ
式中:
第次不稳定性;
δsi——i
第次转速实测值,单位为转每分(r/min);
ni——i
nˉ——转速平均值,单位为转每分(r/min)。
7.7噪声
离心机噪声的A计权发射声压级测量方法如下所述。
a)将离心机放置在坚硬且光滑的反射面(例如地板或测试台面)上,并按制造者规定的方式固定
或安装。墙壁、天花板和任何其他反射体与离心机表面之间的距离不应小于3m。
b)传声器位于距离心机外表面1.00m、地面上方高度(1.550±0.075)m处。测量时,声级计的传
声器应指向离心机。
c)启动离心机前,在步骤b)所述位置上测量并记录背景噪声的A计权声压级。
d)将离心机设置为产生最大噪声的工况,启动离心机,待达到设定转速1min后,在步骤b)所述
位置上测量并记录离心机工作时的A计权声压级。
e)计算在每个传声器位置上测得的离心机工作时的A计权声压级与背景噪声A计权声压级之
8
GB/T30099—2025
差。如该差值小于6dB,则测量结果无效;如大于15dB,无需修正(即取背景噪声修正值
=0dB6dB~15dB3
K1);如在之间,则按公式()计算各传声器位置的背景噪声修正值:
=-10lg(1-10-0.1ΔL)3
K1…………()
式中:
dB
K1——背景噪声修正值,单位为分贝();
ΔL——离心机工作时的A计权声压级与背景噪声A计权声压级之差,单位为分贝(dB)。
f)由各传声器位置测得的离心机工作时的A计权声压级减去相应的背景噪声修正值,即得到
离心机噪声在各传声器位置的A计权发射声压级。取其中最大者为离心机发射噪声声压级
的报告值。
g)当测试地点的海拔高于500m时,应按GB/T17248.2—2018中5.5.2规定的方法,将测得的
离心机发射噪声声压级归一化至标准气象条件。
7.8试液温升
将纯净水放入离心管(瓶)内,用测温仪测量试液的温度,试液温度与环境温度相差±2℃范围内,
离心机在其转头对应的最高额定转速对应最大载荷下,按表1规定的时间运行,停机后,在1min之内
再次用测温仪测量试液的温度,计算两次测量示值之差。
7.9升速时间和降速时间
离心机装上被测转头,配置最大载荷。设置离心机最快升、降速挡位及被测转头的最高额定转速。
启动离心机,测量被测转头转速从零升至最高额定值时所需的时间,其结果应符合表2的要求。当被
测转头在最高额定转速时,按下停止键,测量被测转头从最高额定转速降至零时所需的时间。
7.10振幅
用测振仪在离心机外壳前方、左、右侧面的几何中心进行测量,取其最大值。
7.11制冷
7.11.1制冷时间
将测温仪的传感器安装在离心腔内距温度传感器10mm范围内,装上被测转头,配置最大载荷,
调至该转头的最高额定转速,温度设置为25℃,启动离心机,当达到设置最高额定转速和温度时并稳
定10min后,将温度设置为0℃,同时启动电子秒表开始计时,记录测温仪的温度示值降至4℃时所经
历的时间。
7.11.2温度误差
将测温仪的传感器安装在离心腔内距温度传感器10mm范围内,装上被测转头,配置最大载荷,
调至该转头的最高额定转速,分别设置温度为20℃、15℃和5℃,在每个设置条件下,当测温仪温度示
值达到设置值,稳定运行5min后开始读数,30min内在相同时间间隔下采集不少于10个值,计算读
取值的平均值与设定值的差。
7.12制热
7.12.1制热时间
将测温仪的探头安装在距离心机离心腔内的温度传感器小于10mm处,装上被测转头,配置最大
9
GB/T30099—2025
载荷,调至该转头的最高额定转速,温度设置为25℃,启动离心机,当达到设置最高额定转速和温度时
并稳定10min后,将温度设置为80℃,同时启动电子秒表开始计时,计算测温仪的温度示值升到80℃
时所经历的时间。
7.12.2温度误差
将测温仪的探头安装在距离心机离心腔内的温度传感器小于10mm处,装上被测转头,配置最大
载荷,调至该转头的最高额定转速,分别设置离心腔温度为60℃、70℃和80℃,在每个设置条件下,当
测温仪温度示值达到设置值,稳定运行5min后开始读数,30min内在相同时间间隔下平均采集不少
于10个点,计算读取值的平均值与设定值的差。
7.13安全
7.13.1电气安全
7.13.1.1泄漏电流
按GB4793.1—2007中6.3的规定进行试验。
7.13.1.2介电强度
按GB4793.1—2007中6.8的规定进行试验。
7.13.1.3插头连接设备的保护连接阻抗
按GB4793.1—2007中6.5.1.3的规定进行试验。
7.13.2机械安全
7.13.2.1正常运行位移
在离心机安装位置,用笔沿离心机底部画直线;转头装载80%额定容量的自来水,启动离心机运
行至当前转头的最高额定转速,离心机机脚应无任何偏移。目视检查离心机的位置。
7.13.2.2机体抗倾倒
在离心机的前面或后面、左面或右面底部加装固定点防止侧向移动,使离心机向每一个方向倾斜10°。
在6.11.2.2.2规定的位置施加6.11.2.2.2规定的力。
7.13.2.3非正常运行位移
离心机应在以下组合条件下进行试验:
a)安装方式:按制造商的规定进行安装;
b)转头的选择:在制造商使用说明书规定的转头中,选择动能最大的一个转头进行测试;
c)转头的处理:对其进行预处理,预处理应使其破损状态产生的碎片表面积最大化;
d)转速要求:装置失效时,离心机应在该转头对应的最高额定转速状态;
e)转头容量要求:转头装载额定容量80%,密度大于或等于1.2g/mL带明显显色的试液;
f)不平衡保护:拆除转头不平衡保护功能。
离心机在以上所列组合条件下运行至破损状态,目测离心机位置及可能的移动。
10
GB/T30099—2025
7.13.2.4运动部件的防护
采用直径4mm的销钉检查离心腔顶部。
目测检查离心机机盖及锁紧装置。
目测检查离心机转头组件或驱动系统运动零部件。
7.13.2.5联锁保护
打开离心机机盖,启动离心机,检查驱动装置是否运转;合上机盖,启动离心机,检查机盖是否能够
打开。断开电源,使用工具,检查开锁装置能否开锁。
7.13.2.6不平衡保护
用天平称量不平衡量,模拟不平衡状态,检测不平衡装置是否有效。
7.14电磁兼容性
7.14.1发射
对于所有的试验,被试装置都宜尽可能按实际安装情况进行布置,包括有关的连线和商定的端接。
同时,离心机的发射试验应在7.2要求的试验条件下进行。
离心机的传导发射和辐射发射试验按照GB4824给出的测量方法进行。
离心机的谐波电流发射试验按照GB17625.1给出的测量方法进行。
离心机的电压变化、电压波动和闪烁的限制试验按照GB/T17625.2给出的测量方法进行。
7.14.2抗扰度
对于所有的试验,被试装置都宜尽可能按实际安装情况进行布置,包括有关的连线和商定的端接。
同时,离心机的抗扰度试验应在7.2要求的试验条件下进行。
离心机的抗扰度试验应按照表3中引用标准给出的测量方法进行。
7.15环境适应性
对于所有的试验,被试装置都宜尽可能按实际安装情况进行布置,包括有关的连线和商定的端接。
离心机的环境适应性试验按照GB/T14710给出的测量方法进行。
8检验规则
8.1检验分类
离心机的检验分为:
a)出厂检验;
b)型式检验。
8.2检验项目
出厂检验、型式检验的项目、要求及试验方法的条款号见表6。
11
GB/T30099—2025
表6
序号检验项目要求的条款号试验方法的条款号出厂检验型式检验
1外观6.17.3●●
2结构6.27.4—●
3转速相对误差6.37.5●●
4转速稳定性6.47.6—●
5噪声6.57.7●●
6试液温升a6.67.8●●
7升速时间、降速时间6.77.9●●
8振幅6.87.10
温馨提示
- 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
- 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
- 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
- 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
- 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
- 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
- 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。
最新文档
- T/CPIA 0064-2024光伏封装胶膜表面粗糙度测试方法
- T/TIC 004-2023生态环境监测机构综合能力评价通用规范
- T/FSHW 05-2025道路保洁质量与作业规范
- T/CI 1481-2026农民用水合作组织运行制度设计规程
- T/CAAMTB 191-2024车用70MPa加氢口性能试验方法
- T/CPF 0114-2025温室气体 产品碳足迹量化方法与要求 成型纸盒
- 2025~2026学年安徽省阜阳市临泉县校联考北师大版六年级上册11月期中数学试卷【附答案】
- 2025-2026学年陕西省咸阳市初中政治八年级期末下册自测专项攻坚题附答案
- 人工智能在生物多样性监测中的应用
- Photoshop基础技能考核试题(含详细答案)
- 临床检验标准化实验室完整SOP程序文件
- 2026年大数据工程师笔试试卷(附答案)
- 林下种植林地协议
- 第7课《社会主义好》第一课时(教学课件)
- 面料染色工艺质量控制细则
- 衰弱老年住院患者护理专家共识(2026完整版)
- 2026中国高端装备润滑材料技术门槛与军事应用价值报告
- 2026年抗菌药物合理使用全员培训计划表
- 戒烟健康科普 拥抱美好人生
- 2025山东沂蒙山天蒙旅游区社会招聘2人笔试历年备考题库附带答案详解
- 华北科技学院《电力电子技术》2025-2026学年第一学期期末试卷(A卷)
评论
0/150
提交评论