1、 ECG acquisition system in analog circuit designECG as the heart electrical activity in the performance of body surface, the signal is weak, the spectral range of 0.05 200Hz, voltage amplitude is 0 5mV 1, the signal source impedance of thousands to hundreds of thousands of Europe Europeand there wer
2、e a lot of noise, so the ECG acquisition system can be the rational design of the right of the key components of ECG.ECG measurement conditions is quite complex, in addition to subject, including electromyography, respiratory wave signal, EEG signal interference and other outside interference in viv
3、o has also been 50HZ electricity, baseline drift, electrode contact and interference with other electromagnetic devices in vitroTherefore, the strong noise interference how to effectively suppress a variety of ECG acquisition system will be the key to the design.1 ECG acquisition systemComplete ECG
4、acquisition system includes analog and digital two parts 2, which simulate some of the major to complete the pick up ECG signals, amplification and filtering, the analog portion of the digital part will receive the ECG signal analysis and processing to health carepersonnel arrive at a correct diagno
5、sis, the ECG acquisition system in the analog circuits in the ECG monitoring system plays an important role.ECG acquisition system structure shown in Figure 1.Figure 1. The overall structure of ECG acquisition systemECG acquisition circuit simulationAD conversionLevelof upliftEnlargeFilter circuitPr
6、eamplifier ECGCarried on by the special electrodes in the human ECG signal pick up the first amplified by the initial pre-amplifier, and a variety of interference signals in a certain inhibition into the band-pass filter to filter the ECG signal interference outside the frequency range, and then fil
7、tered by the main amplifier will amplify the signal further to the right range, and then filtered by 50Hz notch filter frequency interference, are simulated ECG signal will be converted into the AD system, analog-digital conversion, converted into digital signalsAfter the central processing unit res
8、ponsible for the subsequent analysis and processing.2 preamplifier circuitPicked up from the body surface ECG generally only a few millivolts, in order to improve its resolution display and processing in the backend, you first need to amplify the signal.ECG acquisition process, the pre-amplifier the
9、 greatest impact on the ECG, ECG to improve the performance of the amplification of preamplifier can not choose too large (generally less than 20), otherwise they will be duelarger interference signal (refer to the electrode polarization voltage), causing the amplifier obstruction.For the ECG, the a
10、cquisition of the signals are differential mode signals, so the amplifier is a differential amplifier circuit structure, the most common is the use of low-noise, high input impedance, high CMRR, high gain and interferencestrong ability in parallel with the differential amplifier circuit, known as th
11、e three op amp instrumentation amplifier, this system uses a common integrated operational amplifier LM324 amplifier to constitute such.LM324 is an integrated operational amplifier 4, as cheap and easy to use, it is widely used in control and signal amplification processed.We are carefully designed
12、and repeated in the laboratory experiments, using LM324 circuit composed of successful amplification of the ECG signal processing, its main technical indicators can meet the requirements.LM324 can either use a single power supply, you can also use the dual power supply, power supply voltage has been
13、 used from the +5 V ± 15 V, and drive with low power consumption, each group of differential gain amplifier can reach 100dB.Through the rational design of peripheral circuits, making the LM324 amplifier as the main device can meet the high magnification, high stability amplification of ECG proc
14、essing requirements.It is worth mentioning that, LM324's low current noise, especially for the application of ECG, it has a very good DC Characteristics: input offset voltage is less than 5mV, the input offset voltage drift is less than 7uV / , the input bias current less than40nA, common mode r
15、ejection ratio of 100dB, and its low power, low power (as low as 3V), and other features to streamline package is battery-powered portable ECG monitoring system an excellent choice.Amplification circuit shown in Figure 2. Preamplifier circuitFigure 2 Figure 2 Design of the amplifier consists of two
16、components, U1C and U1D constitute the first stage of the differential input and output stage, U2A differential ratio for the basic circuit, the total voltage gain Au, is equal to the product of two-stage gain.The first stage uses the same phase as input, a higher input resistance, U1C and U1D op am
17、p selection of the same characteristics, so that their common-mode voltage output voltage and drift are equal, then through the second differential fractional U2A circuit composed ofcan cancel each other out, and the second differential amplifier circuit will be point differential input single ended
18、 output changing to adapt to the needs of ground load.After the two-stage cascade circuit, they are mutually reinforcing, so that combination of this circuit has high input impedance, voltage gain adjustment is convenient, common mode rejection is high and offset drift, and a series of advantages.In
19、 order to improve CMRR and reduce temperature drift effects, and further improve the circuit performance, measuring amplifier with symmetrical structure, ie a few carefully selected external resistor R10, R11, R12, R13, R14, R15, R16 and regulation of RW2, makes the R11= R12, R13 = R14, R16 = R15 +
20、adjusted RW2.So by adjusting the size of an external resistor RW2 can easily change the measurement of amplifier gain.Removal of the body to carry communication mode interference is an effective method is to use the right leg drive circuit 3, in the Figure 2 system, the right leg RL side is not dire
21、ctly grounded, but the amplifier output.Shielded from the driver output common mode voltage detection, it is the reverse by the auxiliary amplifier U0 amplified, and then through the feedback resistor R1 to the right leg, right leg drive thus be the name of that displacement current is not flowing i
22、nto the human bodyland, but the flow of R1 and auxiliary amplifier output of U0.But right leg drive circuit of the AC interference voltage feedback loop, while the AC current flows through the body may have become unsafe, the current limiting resistor is not very small, usually taking more than 1M i
23、n Europe, where R1 is a security protection, between when the patient and to high voltage, the auxiliary amplifier saturation U0, right leg drive circuit does not work, U0 equivalent to grounding, the resistor R1 since then to limit protection.Interference signal sent to the right leg drive amplifie
24、r inverting amplifier, driven right leg electrode RL reached for the depth of interference signal is a negative feedback, which effectively weakens the human body sensing the common mode interference signals.Figure 3The main amplification and filtering circuits3 main amplifier and filter circuitThe
25、main ECG amplification and filter circuit shown in Figure 3.Detection signal as the main interference in the signals between the electrode plates and the polarization voltage, 50Hz frequency interference, the instrument around the internal noise and electromagnetic interference equipment and so on.T
26、o obtain a clear ECG signal stability, filter design is also critical, especially 50Hz band stop filter is particularly important.Collected from the ECG, 200Hz signal is stronger than the interference, and interference signals below 0.05Hz relatively weak, so take the first filter circuit in the low
27、-pass filter out signals below 200Hz, and then take the way of high-passto filter out the polarization voltage and high frequency interference.In the circuit U2B and resistors, capacitors, composed of band-pass filter, while the circuit with high input impedance 4.Filter circuit using RC coupling ci
28、rcuit, high input impedance low output impedance, and input and output between the good isolation.Its role is mainly to pre-amplifier across the DC voltage and DC polarization voltage, coupled ECG.In order to ensure that no distortion of ECG coupled to the next level, must be coupled RC.RC product o
29、f the larger low-frequency response of the amplifier the better, but increase the RC values can not be unlimited, because the R value is limited by the input impedance, C is too large not only large in size and increase in leakage current can also cause drift, but alsocharge and discharge time will
30、be extended loop.ECG body low frequency filter circuit with the RC active filter can effectively avoid the general-purpose integrated circuit operational amplifier as the bandwidth is narrow and not suitable for high frequency range of defects.4 level lift circuitAmplified ECG signal after filtering will be sent to A / D converter analog to digital conversion, and the system chosen AD converter is a single 5V supply the MAX187, so the need to add a filter after the uplift of the circuit level, theECG signal raised to the range of 0 + 5V.Circuit design is very simple, shown in Figu
温馨提示
- 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
- 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
- 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
- 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
- 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
- 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
- 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。
最新文档
- 家长对孩子的养成教育
- 重阳节活动的策划方案
- 专职安全生产管理人员综合类C证模拟题含答案(附解析)
- 导游证基础模拟练习题与答案(附解析)
- 2024年12月继电保护员-中级工习题库(含参考答案解析)
- 绿化施工环境保护措施考核试卷
- 班主任工作技能课件
- 矿产勘查地球物理方法应用考核试卷
- 2025年IC卡鉴别机项目建议书
- 《企业质量管理策略与应用》课件
- 工程项目全过程咨询管理
- 【MOOC】温病学-河南中医药大学 中国大学慕课MOOC答案
- 【MOOC】美术鉴赏-河南理工大学 中国大学慕课MOOC答案
- 《汽车用改性聚丙烯车门外板》
- 栏杆拆除及更换施工方案
- 《城市轨道交通车辆构造》 课件 5.1 城市轨道交通车辆车门总体认知
- 腐蚀速率实时预测
- 2024秋期国家开放大学专科《会计信息系统(电算化会计)》一平台在线形考(形考任务一至四)试题及答案
- 送快递劳务承揽协议书
- 2024年安徽安庆市交通控股集团有限公司招聘笔试冲刺题(带答案解析)
- 贷款中介服务合同
评论
0/150
提交评论