1、New proposed methodology for specific heat capacity determination of materials,for thermal energy storage (TES) by DSC,1,峰谷书屋,Abstract,1.Introduction,2.Materials,3.Methodology,4.Results,5.Conlusions,2,峰谷书屋,abstract,This study presents a methodology to determine the specific heat capacity (Cp) of mat

2、erials for thermal energy storage (TES) by DSC.Three different DSC methods are considered to be applied in the methodology, and are explained and compared in this study in order to select the most proper one for Cp determination. To perform this study, the Cp of three materials commonly used in sens

3、ible TES systems, slate, water, and potassium nitrate (KNO3), is determined.,3,峰谷书屋,一、introduction,lack of fossil energetic resources its price need to decrease CO2 emissions lead to dedevelop new energy systems able to take profit of renewable energy and be environmentally friendly and less expensi

4、ve,TES systems are a possible solution Solar energy abundant clean and easy accessible. problem:as the hours of maximum energy demand match with the hours of no solar irradiation,with the aim to fill this existing energy supply gap, TES systems are presented as the solution to store the energy and u

5、se it anytime,4,峰谷书屋,TES technologies sensible heat storage the process by which the heat is accumulated due to the increase of the material temperature without experimenting structural changes, no phase change latent heat storage phase change , thermochemical energy material selection A high storag

6、e capacity Therefore,need a high specific heat capacity.,(1),5,峰谷书屋,定义:在程序控制温度下,测量物质与参比物之间的能量差随温度变化关系的一种技术。 DSC曲线峰包围的面积正比于热焓的变化 可以测定多种热力学和动力学参数,例如比热容、反应热、转变热、相图、反应速率、结晶速率、高聚物结晶度、样品纯度等。该法使用温度范围宽(-175725)、分辨率高、试样用量少。适用于无机物、有机化合物及药物分析。,差示扫描量热法(DSC) (Differential Scanning Calorimetry),6,峰谷书屋,采用试样和一标准物质在

7、相同条件下进行扫描,然后量出二者的纵坐标进行计算。要求标准物质在所测温度范围内没有发生物理化学变化,且比热容已知。常用的标准物质是蓝宝石。具体作法是在DSC内先用两个空样品皿,以一定的升温速度作一条基线,然后放入蓝宝石样品在相同条件下作一条DSC曲线,最后在同样条件下作未知试样的DSC曲线,如图,7,峰谷书屋,three DSC methods for Cp measurement The dynamic method has traditionally been used to measure PCM latent heat but also for Cp measurements appl

8、ying high heating rates in the temperature range of study The areas method is based on the Cp value, thus, how much heat flux is needed to heat up a material for temperature increase, and consists of consecutive isothermal stages differing 1 with no heating segments amid The isostep method is an int

9、ermediate between these two, as it is a succession of dynamic methods applied to heat up the material just 1 between isothermal segments. This method has been tested in glass transition pharmaceutical studies with good results obtained , but it has not been used for TES materials Cp determination, t

10、hereby its selection for this study,8,峰谷书屋,二、Materials,To ensure the performance of the methodology in a representative variety of material phase forms and chemical structures, three different materials widely used in sensible heat storage systems have been chosen to perform this study. Water. Its t

11、hermal properties are well known, and therefore, these values can be used as a reference to see the approach on the measurements each method has. Commercial Bi-distillated water from Panreac has been used to perform the experiments Slate. It is a widely used construction material and its performance

12、 depends on its specific heat capacity, therefore authors find it an interesting material to be tested. The slate samples used in this study were taken from a quarry in the Catalan Pyrenees. Potassium nitrate. It is an inorganic salt mainly used as molten salt in concentrated solar power plants (CSP

13、 Plants). It has high melting point (320) and, as all inorganic salts, a complex chemical structure, and that is why it is also found to be useful for this papers purpose. It is important to remark that this material selection responds to the need to test the methodology and not to tie it to a concr

14、ete material type or a specific TES application. Therefore, materials used in different TES systems and with which authors commonly work have been selected.,9,峰谷书屋,determined by differential scanning calorimetry (DSC) Standard 40 uL aluminium crucibles weigh with a analytical balance with a precisio

15、n of 0.01 mg. The procedure to determine the specific heat capacity of a material consists of three different measurements, all done under the same conditions, thus, using the same DSC method: Blank measurement. It is necessary to run an experiment with an empty crucible to measure the heat flux tha

16、t corresponds to the crucible material in order to subtract this signal and consider only the sapphire and material sample ones. Sapphire measurement. Sapphire is the material used as reference as its specific heat capacity is well known at every temperature and its signal is hugely stable over temp

17、erature, data necessary for the materials Cp calculation. Material measurement. Needed to determine the Cp value of the material under study.,三、Methodology,mW,K/S,(2),10,峰谷书屋,The sensitivity of the analysis is tied to the heating rate and the sample mass. literature studies found high heating rates

18、of about 1020 K/min are desired Eq. (2) shows, minor sample masses are needed. The Cp was measured between 15 and 16 , 2526 and 35 36 for all the materials under study. These measurement ranges were selected in order to ensure that water was in liquid state and to be able to take the certain known w

19、ater 4.18 J/gCp value at 25 as a reference to evaluate each method performance.,3.1. DSC measurement methods 3.1.1.Dynamic method It is a temperature controlled method that achieves the thermal equilibrium before and after a controlled heating segment ,it consists of three segments. 1015 min to homo

20、genize the material temperature high heating rate another isothermal stage again for 1015 min the lower and higher temperatures of the method can be 150 ,(3),11,峰谷书屋,Similar to the previous explained dynamic method, the isostep method This method uses Eq. (3) to calculate the Cp of the material alon

21、g each heating segment.,3.1.2.isostep method,12,峰谷书屋,3.1.3.Areas method The areas method consists of consecutive isothermal segments with no heating stages amid.Each of the peaks corresponds to every temperature step of the method. The temperature difference between isotherms is of 1 .,where As J/g

22、is the integrated peak area for the sapphire curve,13,峰谷书屋,In this study, results are given with an error of 3%, including both equipment and calculation errors. The dynamic method gives a continuous signal along the whole temperature study range, as shown in red Three different continuous signals a

23、re obtained with the isostep method, one for each heating segment for the 1516 C, 25 26 C and 3536 C steps. However, as the measurement temperature range is of just 1 C and the material is heated at a really fast rate,四、results,the three signals are really steep and no clear tendency can be seen in

24、the results, fact that already shows the low sensitivity this method has for Cp calculation. Nevertheless, and in order to compare the methods, an average value of the most constant parts of these three signals is presented in the graphics. Contrary, the areas method provides just three points, one

25、for each temperature increment between the isothermal segments, hence, one at 16 C, one at 26 C and one at 36 C.,14,峰谷书屋,4.1water,areas method gets closer values to the theoreticai values The dynamic method inconsistent values are found in the begining and the end,that indicates sensitivity problems

26、 in these two measurement parts. the isostep method are the average of the most constant parts of the obtained signal,being the standard deviation for water of 0.08 J/g .,The sensitivity problems. Due to the abrupt temperature change on both initial and end points, the DSC temperature sensor cannot

27、react fast enough to read the real temperature.This fact has greater significance in the isostep method results due to the higher relation between heating rate and temperature increase.,15,峰谷书屋,results show that the areas method is the method that provides Cp close to the theoretical range dynamic m

28、ethod the isostep method, with a standard deviation of 0.03 J/g .most inconsistent,4.2.slate,16,峰谷书屋,4.3. potassium nitrate,same condition same reason,17,峰谷书屋,both dynamic and isostep method measurements are limited by the high heating rates applied, that low down the sensitivity of the analyses and increase the error of the results. comparing the three material results analysis, at t

温馨提示

  • 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
  • 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
  • 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
  • 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
  • 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
  • 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
  • 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。

评论

0/150

提交评论