1、Vascular Protection Training,Advanced Clinical Training,1,学习交流PPT,The Product,Features and Intended Benefits,2,学习交流PPT,FilterWire EZ Embolic Protection System,The FilterWire EZ System is BSCs second generation embolic protection filter. It incorporates a number of design changes from the FilterWire
2、EX System (First Generation System) intended to improve the effectiveness and ease of use. Europe September 2003 commercialization for SVG, native coronary, peripheral and carotid intervention Dominant market share (PV and CV) despite many other embolic protection filter competitors US August 2004 c
3、ommercialization for SVG intervention Increased market share and market penetration December 2006 510(k) approved for carotid indication,3,学习交流PPT,FilterWire EZSystem Components,FilterWire EZ 5.5mm System Protection Wire EZ Delivery Sheath EZ Retrieval Sheath Accessories(peel-away introducer, guide
4、wire torquer, valve dilator tool) FilterWire EZ Bent Tip Retrieval Sheath (available separately),4,学习交流PPT,FilterWire EZ SystemAccessories,Peel-away Introducer Sheath Used to prevent damage to the spring tip of the protection wire during system insertion into a rotating or fixed hemostasis valve.,Gu
5、ide Wire Torquer Intended to enhance the operators grip on the protection wire for steerability during delivery and to provide leverage during filter deployment and capture.,Valve Dilator Tool Used with a fixed hemostasis valve to minimize damage to the filter bag upon removal through the valve.,5,学
6、习交流PPT,FilterWire EZ System,PEBAX is a registered trademark of Atofina Corporation,Atraumatic 3 cm, platinum, spring coil tip with floppy flexibility Radiopaque to enhance visualization Shapeable to enhance steerability 6.0 mm PEBAX Nosecone Flexible transition from the spring tip to the delivery sh
7、eath for excellent overall deliverability Provides a pro that is designed to reduce the risk of debris embolization during system delivery,6,学习交流PPT,FilterWire EZ System,Polyurethane filter membrane with 110 micron pores Designed to captures debris and reduce the risk of embolic related events Inten
8、ded to permit continuous blood flow for patient safety and comfort Intended to permit continuous blood flow that allows contrast injections for visualization during the intervention,7,学习交流PPT,FilterWire EZ System,Conforming, nitinol filter loop with platinum coil Highly visible to guide filter place
9、ment and to ensure good vessel wall apposition One device that accommodates vessels with landing zone diameters ranging from 3.5mm to 5.5mm minimizes sizing guesswork and inventory needs Designed to provide continuous wall apposition around the circumference of the vessel lumen to minimize the oppor
10、tunity for debris to collect on the outside of the filter Automatically adjusts to changes in the vessel aperture,8,学习交流PPT,FilterWire EZ System,Filter loop suspension arm allows the filter loop to move independently from the wire for good vessel wall apposition in both straight and curved anatomy S
11、implifies the landing zone requirements by allowing use in both straight and curved vessels No need to take 2 angio views to assess filter apposition,9,学习交流PPT,FilterWire EZ System,Spinner tube underlying filter assembly Allows the wire to rotate independently from the delivery sheath for excellent
12、system steerability .032” OD Catheter stop Provides a barrier designed to prevent therapeutic devices from collapsing the filter frame Lightly visible to contribute to visualization for device positioning*,* Note: A 0.013” OD part 5mm proximal to the catheter stop may also be visible with some equip
13、ment. This part keeps the spinner tube in place, but lets catheter tips pass over.,10,学习交流PPT,FilterWire EZ System,0.014” Stainless steel wire body with PTFE coating Available in 190cm and 300cm wire lengths 190cm wire length is compatible with BSCs AddWire GuideWire Silicone coated on distal 15 cm
14、of device, delivery, and retrieval sheath. Designed to reduce the amount of force required to load the filter into the delivery sheath and deploy in the vessel,11,学习交流PPT,EZ Delivery Sheath,150 cm PEBAX Resin sheath with 100cm, peel-away exchange length The distal sheath changes color from blue to c
15、lear to serve as a warning that the non peel-away section at the end of the sheath is approaching 1.1mm (0.042” or 3.2F) Lesion crossing profile Distal end has a PTFE inner liner Internal stop loading feature alerts the operator that the filter is fully sheathed,12,学习交流PPT,EZ Retrieval Sheath,150cm
16、PEBAX Resin sheath with a 20cm long exchange length Soft distal tip Designed to facilitate the capability to cross deployed stents 1.4mm (0.056” or 4.3F) lesion crossing profile Internal stop capture feature Alerts the operator that the filter is fully captured,13,学习交流PPT,EZ Bent Tip Retrieval Sheat
17、h (sold separately),150cm PEBAX Resin sheath with a 20cm exchange length Bend at distal tip The bend is designed to aid in crossing stent struts that can complicate filter retrieval 1.3mm (0.052” or 4.0F) lesion crossing profile Internal stop capture feature Alerts the operator that the filter is fu
18、lly captured,14,学习交流PPT,FilterWire EZ Embolic Protection System,Step-by-Step Instructions Product Demonstration,Please refer to the FilterWire EZ Embolic Protection System DFU for complete instructions for use.,15,学习交流PPT,FilterWire EZ Embolic Protection System DFU available at,FilterWire EZ SystemD
19、FU Card,16,学习交流PPT,FilterWire EZ Embolic Protection SystemPre-Procedure,Review the patients angiogram and determine if the vessel meets the requirements for safe and effective use of the FilterWire EZ System: The apex of the loop can be deployed a minimum of 3.0cm from the distal edge of the lesion.
20、 The vessel diameter at the desired location of the deployed filter loop is 3.5mm to 5.5mm. (2.25mm to 3.5mm for EZ Small),3.5mm 5.5mm vessel diameter,17,学习交流PPT,FilterWire EZ Embolic Protection SystemPre-Procedure (continued),Prior to using the FilterWire EZ System, anticoagulate the patient to ach
21、ieve an ACT of 275 seconds for carotid applications and 300 seconds for SVG applications (200 seconds if GP IIb/IIIa inhibitors are used). In the absence of ACT measurement, administer an appropriate weight based bolus of heparin (approximately 125 units/kg).,18,学习交流PPT,FilterWire EZ Embolic Protect
22、ion SystemPre-Procedure (continued),RCA graft - most commonly used guide is the multipurpose. Marginal graft - Hockey stick guide is first choice followed by left Amplatz,LAD or Diagonal graft - most commonly used guides are the left Amplatz and Hockey stick.,Taken from Marco et.al.“Stenting of Dege
23、nerated Saphenous Vein Grafts” May 2001 Euro PCR,Based on vessel take-off, physician should consider which guide catheter shape will provide the best access and support.,19,学习交流PPT,FilterWire EZ Embolic Protection System PREP: Delivery System,1.Open the pouch using sterile handling procedure and pla
24、ce the packaging coils and accessory kit into the sterile field. 2.Unclip and remove the yellow housing that protects the filter. Note: The protection wire is pre-loaded into the delivery sheath. 3.Detach the pre-loadedprotection wire from the retaining clip, advance the delivery sheath, then grasp
25、theclear section of the sheathand removesheathed wire from the coil.,FilterWire EZ System Product Packaging,20,学习交流PPT,4.Sheath the protection wire and filter by retracting the device into the EZ Delivery Sheath while the filter is immersed in heparinized saline.NOTE: Hold the Delivery sheath about
26、1cm proximal to the dark section. NOTE: Keep the filter unsheathed until immediately prior to use.,FilterWire EZ Embolic Protection System PREP: Delivery System (continued),3.Attach the wire torquer onto the proximal end of the protection wire.,Guide Wire Torquer,21,学习交流PPT,FilterWire EZ Embolic Pro
27、tection SystemDelivery,2.Insert the FilterWire EZ System and peel-away introducer assembly into a guide catheter/ sheath. Once the FilterWire EZ System is inserted into the guiding catheter or guiding sheath, remove the peel-away introducer. NOTE: Use of a guiding catheter or a guiding sheath with a
28、 1.91mm (0.075”) or larger rotating hemostasis valve is recommended when performing a procedure with the FilterWire EZ System. The use of a sheath /guide catheter with a fixed hemostasis valve may cause the filter membrane to tear at the hemostasis valve upon removal.,1.Place the peel-away guide wir
29、e introducer over the tip of the delivery system (the spring tip of the system can be shaped as needed).,Peel-away Introducer Sheath,22,学习交流PPT,FilterWire EZ Embolic Protection SystemDelivery (continued),3.To advance the FilterWire EZ System through the guiding catheter or guiding sheath, use a one-
30、handed technique advancing both the delivery sheath and protection wire together; this will prevent accidental deployment of the filter. 4.To steer the FilterWire EZ System to the target landing zone, use a two-handed technique, torquing the protection wire with one hand and advancing the delivery s
31、heath with the other hand. NOTE: For carotid interventions, having the patient turn their head to the left or right will change the orientation of the vessel and may facilitate device passage. NOTE: Consider using the buddy wire technique if the lesion is difficult to pass. Retract the buddy wire af
32、ter the lesion has been crossed by the FilterWire EZ System (If other efforts are unsuccessful, pre-dilatation can be used.),23,学习交流PPT,FilterWire EZ Embolic Protection SystemDelivery (continued),5.Advance the FilterWire EZ System across the lesion until the apex of the filter loop can be deployed a
33、 minimum of 3.0cm distal to the lesion. NOTE: If an interventional device with a tip-to-shoulder length of greater than 10.0mm is to be used, a longer vessel length may be required to place the FilterWire EZ System in order to avoid contact between the filters catheter stop and the tip of the interv
34、entional device.,24,学习交流PPT,Once the protection wire is in place past the lesion, slide the torquer tip up to the hemostatic valve and tighten firmly. Deploy the filter by holding the protection wire with the torquer, then retracting and completely removing the delivery sheath. Be sure to keep the t
35、orquer in place as you peel-away the delivery sheath in a straight motion until the 3cm over-the-wire section is reached (see photo below).,FilterWire EZ Embolic Protection SystemFilter Deployment,Photograph taken by Boston Scientific Corporation,25,学习交流PPT,4.Once the delivery sheath is completely r
36、emoved, inject contrast and verify the device is in the proper position and there is adequate flow.NOTE: If repositioning is required, the EZ Retrieval Sheath should be used. Retract the protection wire and filter loop into the sheath, reposition the filter and redeploy. 5.Use the protection wire li
37、ke a traditional guide wire to track catheters and stent delivery systems to the target treatment area.,FilterWire EZ Embolic Protection SystemFilter Deployment (continued),26,学习交流PPT,1.Inject contrast frequently and always after any intervention or exchange, to visually verify that blood flow is no
38、t obstructed and that filter loop/vessel apposition to the vessel wall is maintained.CAUTION: If a filter becomes occluded during the procedure, remove it and deploy a second filter if further intervention is required. 2.Complete the procedure making sure the protection wire remains in the proper po
39、sition distal to the lesion (monitor fluoroscopy). Note: Failure to stabilize the protection wire could lead to inadvertent movement of the filter, resulting in protection wire entanglement and/or delay in the procedure.,FilterWire EZ Embolic Protection SystemDuring the Procedure,27,学习交流PPT,FilterWi
40、re EZ Embolic Protection SystemFilter Capture and Retrieval,1.Remove the EZ Retrieval Sheath from its packaging coil and flush with heparinized saline. 2.Advance the retrieval sheath over the protection wire past any deployed stents up to the proximal end of the filter loop.,Note: If the EZ Retrieva
41、l Sheath is not able to pass the deployed stent, be sure the stent is well apposed and consider post-dilitation. If the Retrieval Sheath is still unable to cross, try the EZ Bent Tip Retrieval Sheath (available separately).,EZ Bent Tip Retrieval Sheath Packaging,28,学习交流PPT,FilterWire EZ Embolic Prot
42、ection SystemFilter Capture and Retrieval (continued),3.Gently and slowly retract the protection wire and filter loop back into the retrieval sheath until resistance is felt. 4.Observe the retrieval procedure under fluoroscopic imaging. Note: The distal edge of the collapsed filter loop should align
43、 with (see figure B), or be proximal to (figure A), the retrieval sheath marker band.,Retrieval Sheath Marker Band,A,B,29,学习交流PPT,FilterWire EZ Embolic Protection SystemFilter Capture and Retrieval (continued),NOTE: If resistance is felt before the distal edge of the collapsed filter loop is aligned
44、 with the marker band, it is possible that there is a large amount of embolic debris in the filter. Carefully retract the protection wire and EZ Retrieval Sheath as a system. The distal part of the filter bag will be outside of the sheath during retrieval in this case; however, the filter loop must
45、be pulled fully into the sheath.,Fully Retracted Filter,Large Debris Load in Filter,30,学习交流PPT,FilterWire EZ Embolic Protection SystemFilter Capture and Retrieval (continued),5.If any resistance is met during retraction of the sheathed FilterWire EZ System, slightly advance the sheathed FilterWire E
46、Z System and rotate the sheath before continuing to retract. 6.Slowly and carefully retract the entire system until the tip of the retrieval sheath is adjacent to the tip of the guiding catheter or guiding sheath. 7.If there is any resistance, retract the guiding catheter and FilterWire EZ System to
47、gether. This is done in case there is a large amount of embolic material captured in the filter that may not fit inside the guiding catheter.,31,学习交流PPT,FilterWire EZ Embolic Protection SystemFilter Capture and Retrieval (continued),8. Remove the FilterWire EZ System from the patient. NOTE: When usi
48、ng a rotating hemostasis valve, ensure that the hemostasis valve is fully opened prior to FilterWire EZ System removal. CAUTION: If a guide sheath with a fixed hemostasis valve is used, the valve dilator tool is required for FilterWire EZ System removal. Removing the FilterWire EZ System through a f
49、ixed hemostasis valve without the use of a valve dilator may cause the filter membrane to tear. 9. If a fixed hemostasis valve is used, place the valve dilator (if provided in the accessory kit) on the shaft of the retrieval sheath and slide the tool into the valve. Then remove the FilterWire EZ Sys
50、tem from the patient through the valve dilator tool and remove the dilator.,32,学习交流PPT,Stent Entanglement,Potential for a distal embolic protection filter to entangle with a stent is inherent to interventional procedures using these devices.,SAFE PRACTICE FOR DISTAL PROTECTION FILTERS: Achieve good
51、guide catheter or guide sheath support Ensure sufficient landing zone Keep the filter in place during deployment Minimize wire movement during the procedure Achieve full stent apposition,33,学习交流PPT,FilterWire EZ Embolic Protection SystemTips for Use,34,学习交流PPT,Competition,RX ACCUNET (Abbott) Angiogu
52、ard (Cordis) Emboshield System (Abbott) Spider (ev3),35,学习交流PPT,CompetitionFixed Wire Filters,Sheathed Fixed Wire Devices: FilterWire EZ System, RX Accunet, Characteristics Filter attached to a proprietary wire Delivery sheath required,Photo by Boston Scientific,36,学习交流PPT,CompetitionWire of Choice
53、Filters,Wire of Choice Device: SpiderFX Device, Emboshield Characteristics: Use any .014” wire to gain access Deliver sheathed filter and proprietary guide wire over the wire of choice, then remove wire of choice Wire can move slightly (8mm of longitudinal “play”) without moving filter CANNOT remove
54、/replace deployed filter while maintaining access,SpideRX ,37,学习交流PPT,CompetitionOcclusive Devices,Distal Occlusion Device: Medtronic GuardWire System Characteristics: Proprietary wire with balloon attached Low crossing profile Balloon inflated distal to lesion Short landing zone Aspirate to remove
55、emboli Proximal Occlusion Devices: St. Jude Proxis System Characteristics Occlusive protection prior to crossing lesion Proxis stops flow and then aspirates to remove emboli,Proxis,GuardWire,38,学习交流PPT,Image by Boston Scientific Corp.,Abbott Vascular Devices,RX ACCUNET Embolic Protection System,39,学
56、习交流PPT,CompetitionRX ACCUNET System,Delivery,Retrieval,Filter Bag,Delivery method: Sheathed wire,Images by Boston Scientific Corp.,40,学习交流PPT,CompetitionRX ACCUNET System,Technical specs excerpted from RX ACCUNET System Directions for Use and other literature published by Abbott Vascular at as of Ma
57、rch 10, 2008,41,学习交流PPT,CompetitionRX ACCUNET System,Information excerpted from FilterWire EZ System and RX ACCUNET System Directions for Use and other literature published by Abbott Vascular at as of March 10, 2008,42,学习交流PPT,Excellent Deliverability Guide wire and delivery sheath enter vessel as o
58、ne unit Highly flexible to navigate tortuous anatomy Low pro smooth transition from tip to filter basket .014” wire is torqued independent from filter basket Accurate Deployment Rapid exchange unsheathing for stability and simplicity Peel away delivery sheath allows for controlled deployment Capture
59、 and Control for Optimal Protection Highly flexible filter basket designed to conform to tortuosity Designed to capture high volume while allowing adequate blood flow Concentric design with guide wire centered in filter Four platinum radiopaque markers surround filter for visualization of wall apposition Secure Filter Retrieval Choice of two rapid exchange recovery catheters designed specifically for the challenges of filter recovery. RX ACCUNET Recovery Catheter Shapeable distal tip steers catheter from stent struts during advancement Filter basket secured with all ca
温馨提示
- 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
- 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
- 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
- 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
- 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
- 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
- 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。
最新文档
- 金属加工行业市场供需研究及投资发展评估规划分析研究报告
- 2026四川内江市隆昌市人民政府金鹅街道办事处招聘11人笔试参考题库及答案详解
- 氯化钙产业行业市场现状供需分析及投资评估规划分析研究报告
- 2026重庆沙坪坝陈家桥街道招聘笔试备考试题及答案详解
- 2026重庆市体育局直属事业单位考核招聘运动员59人笔试备考试题及答案详解
- 2026五一农场医院招聘(2人)笔试参考试题及答案详解
- 能源开采业市场供需分析及投资评估规划分析研究报告
- 2026山东大学法学院非事业编制人员招聘1人考试备考试题及答案详解
- 医学实验室间检测结果互认机制构建与质量控制研究
- 中国螺纹钢市场前景调研与未来发展趋势预判研究报告
- 事业单位汽车驾驶员晋级初级工中级工高级工技师考试题库及答案
- 2026宁麓置地(宁波)有限公司招聘11人考试参考题库及答案详解
- 2026年云南省高考历史试卷(含答案及解析)
- 2026年永修县招聘交通协管人员23人笔试备考试题及答案详解
- 2026汽车驾驶员(技师)考试题库及答案
- DB32/T+5381-2026+万兆数智园区建设导则
- 2026年石家庄工商职业学院教师招聘考试备考试题及答案解析
- 雨课堂学堂在线学堂云《新中国史(扬州)》单元测试考核答案
- 广州市中考化学知识点归纳总结
- 中试总结报告模板
- 矿棉吸音板吊顶专项施工方案
评论
0/150
提交评论