1、MRI Safety,Jerry Allison, Ph.D.,Caveat,This lecture emphasizes MRI safety for high field (1.5T and 3.0T) superconducting magnets such as those used at MCGHI.,MRI Safety,Projectile Accidents Foreign Materials in the Body Other Stuff,Projectile Accidents,The MRI magnets are ALWAYS on (24 hours/day, 36
2、5 days/year) There is a STRONG fringe magnetic field around the magnets The fringe magnetic field is confined to the scan room,4,Projectile Accidents,The fringe magnetic field can cause ferrous metal objects to fly into the bore of the magnet with great force and speed Ferrous metal objects contain
3、iron (steel) The closer you are to the bore of the magnet, the stronger the fringe field,5,As you approach the magnet, the fringe magnetic field gets STRONGER,6,7,8,9,10,11,My Favorite Horror Story Heres one I heard from an Oxford Magnets engineer which was later independently verified by a technolo
4、gist who had been hired to work at this site. A brand new magnet had just finished being installed into a brand new building. All of the acceptance testing had been completed and the magnet was to be turned over to the customer the very next day. There was only one minor problem to be dealt with fir
5、st. One of the sprinklers in the scan room had a tiny leak. A welder was brought in to fix the leak, but somebody forgot to tell him that the magnet was at field. So. in walks this welder with his acetylene torch system. His tank flies into the magnet, the valve breaks off, sparks and catches fire.
6、Since he was in there to fix a leak in the sprinkler system, it had been turned off first. The brand new building burned to the ground!,Projectile Accidents,The fringe magnetic field can be deadly Valhalla, N.Y. Westchester County Medical Center July 27, 2001 Six year old Michael Colombini was fatal
7、ly injured “when an oxygen canister was turned into a guided missile by the powerful MRI magnet”,12,Fatal Accident Details,Michael Colombini was in the bore of a GE Signa An anesthesiologist was in the scan room with the patient The built-in oxygen delivery system failed A nurse heard the anesthesio
8、logist calling for oxygen,13,Fatal Accident Details,The nurse delivered a STEEL oxygen cylinder to the anesthesiologist thinking it was safe (ALUMINUM) Within 3 to 6 feet of the magnet, the STEEL cylinder flew into the bore, fatally injuring the patient,14,Foreign Materials in the Body,Foreign mater
9、ial in the body is a concern in MRI Foreign material is any material in the body other than the flesh and bones that we are born with Wide variety of foreign material is intentionally and unintentionally present in patients, health care providers and others,15,Foreign Materials in the Body,The prese
10、nce of certain foreign materials in the body can be fatal in an MRI magnet Deaths have occurred due to the presence of aneurysm clips and cardiac pacemakers in patients undergoing MRI Sight has been lost by movement of ferrous metal in the eye resulting in vitreous hemorrhage (2.0 mm x 3.5 mm iron f
11、ragment 0.35 T) Prostheses are usually securely anchored to bone and are usually non-ferrous.,16,Foreign Materials in the Body,Fringe magnetic fields stronger than 5 Gauss can affect cardiac pacemakers. To protect patients and others having pacemakers there is a 5 Gauss Exclusion Zone around MRI mag
12、nets. At MCGHI, the Exclusion Zones are the MRI scan rooms,17,Fringe Diagram of Active Shield1.5 T*,*Siemens Symphony,18,16 6”,26 4”,5 Gauss,Safety (continued),-At least two deaths have resulted from MRI Scans on patients with pacemakers.,19,Foreign Materials in the Body,Heating of metals and implan
13、ts has not been significant (even for large prostheses) There are resources for assessing safety of foreign materials in MRI F.G. Shellock. Reference manual for Magnetic Resonance Safety: 2003 Edition Salt Lake City: Amirsys Inc. 2003.,20,Signage,FDA Guidance for the Submission Of Premarket Notifica
14、tions for Magnetic Resonance Diagnostic Devices states: The controlled access area should be labeled Danger - High Magnetic Field at all entries. The term warning does not convey the importance of a situation that may not only be potentially hazardous, but has been responsible for serious injuries a
15、nd deaths.,21,22,Other Safety Issues Static Magnetic Fields Time Varying Magnetic Fields RF Magnetic Fields Auditory Safety FDA Safety Guidelines Contrast Agents,M R I Physics Course,23,Other Safety Issues Psychological Distress Cryogen Safety Experiences at Higher Magnetic Fields Pregnancy,M R I Ph
16、ysics Course,24,Safety,Why worry about biological and safety issues? -The patient is exposed to: Static magnetic field B0: 3.0 T Time varying magnetic fields (gradients) RF magnetic fields (B1) Acoustical noise Psychological distress (claustrophobia) Contrast,25,Safety (continued),-The patient is al
17、so exposed to: Metal implant accident (foreign materials in the body) Cryogen accident Projectile acident,26,Safety (continued),The technologist, health care providers and other individuals are exposed to: Static magnetic fields: few Gauss to 3.0 T Cryogen accident Projectile accident,27,Safety (con
18、tinued),Biological effects of static magnetic fields -Augmentation of the T wave during peak flow in the aortic arch -No evidence of mutagenic effects (in vitro or in vivo) -No evidence of an effect on organ development in vivo -No pronounced threshold -No trends suggestive of a dose response curve,
19、28,Safety (continued),Biological effects of static magnetic fields “In general, there is no conclusive evidence for irreversible or significant biological effects related to acute, short-term exposure of humans to static magnetic field strengths up to 2.0 T.”,29,Safety (continued),Biological effects
20、 of time varying (gradient) magnetic fields -Faradays Law: Induction of electric fields and currents in tissues (depends on tissue conductivity, gradient strength, risetime, waveshape, polarity, repetition time TR and duration of the scan) -Eddy current formation,30,Safety (continued),Biological eff
21、ects of time varying (gradient) magnetic fields -Nerve stimulation: electrical stimulation of peripheral and cutaneous nerves should precede any stimulation of ventricular fibrillation. Guidelines for time varying magnetic fields are designed to prevent any perceptible nerve stimulation ( with a saf
22、ety factor of times three) -Magnetophosphenes: perceptible light flashes caused by electrical stimulation of the retina (not experienced in imagers having field strengths less than 4.0 T),31,Safety (continued),Biological effects of time varying (gradient) magnetic fields “guidelines are believed to
23、provide a wide margin of safety with respect to exposure to gradient magnetic fields during MRInewer “fast scanning” techniques that may require more complex applications of gradient magnetic fields may easily exceed these recommended levels and must be thoroughly evaluated for potential health haza
24、rds.” -Time varying magnetic fields are used to stimulate bone healing,32,Stimulation Threshold Levels,Mean Cardiac Stimulation Threshold Mean Respiratory Stimulation Threshold Mean Painful Nerve Stimulation Threshold Mean Peripheral Nerve Stimulation Threshold,T/S = Tesla per second 0 - 20 T/s = cl
25、inical mode 20 T/s = First or Second Controlled Modes* * Limited by IRB or Human Ethical Committee,33,3600 T/s,900 T/s,90 T/s,60 T/s,45 T/s Typical Operating Range,20 T/s,0 T/s,MR Safety Guide 2003 by General Electric Company, Inc.,Due to the rapid rate of change of the magnetic fields (dB/dt) used
26、during some EPI scans, a percentage of patients may experience a tingling or touch sensation. This means a percentage of patients may experience Peripheral Nerve Stimulation (PNS) at 45 T/s. If this sensation is bothersome or uncomfortable to the patient, stop the scan. Change to a non-EPI pulse seq
27、uence to continue scanning the patient.,34,MR Safety Guide 2003 by General Electric Company, Inc.,Safety (continued),Biological effects of RF magnetic fields -RF magnetic fields are oscillating magnetic fields -These fields produce heating of tissue -Energy deposited (degree of heating) is measured
28、by the Specific Absorption Rate (SAR) in units of Watts/kilogram,35,Safety (continued),Biological effects of RF magnetic fields - SAR depends upon flip angle, TR, tissue density, RF pulse duration and frequency, RF coil radius, tissue conductivity and static magnetic field. -Guidelines restrict rise
29、 in core temperature to 1 oC FDA: 0.4 W/kg whole body 3.2 W/kg head 8.0 W/kg peak,36,Safety (continued),Biological effects of RF magnetic fields -Similar to ultrasound since as long as the energy deposited is below established limits, minimal heating of tissues occurs. -Extra precautions are necessa
30、ry for patients with compromised thermoregulation: fever, diabetes, obesity, cardiovascular disease, old age, medicated ( and Ca channel blockers), anesthesia,37,Safety (continued),Biological effects of RF magnetic fields -Thermal burns -Electrical conductors can absorb energy and become extremely h
31、ot -Avoid having any cable wire or lead in contact with the patient -Surface coil cables can be placed in sleeves to prevent burns,38,Safety (continued),Auditory safety Activation of gradient magnetic fields produces significant vibrations in the gradient coils. MRI acoustical noise has been shown t
32、o produce reversible hearing impairment and could potentially produce permanent damage. Hearing protection is recommended for all patients undergoing an MRI procedure on a high-field MRI system (1.5T and 3.0T). Noise attenuating ear-plugs or head phones are routinely used in MRI,39,Safety (continued
33、),-FDA Safety Guidelines for MR Devices -Acoustic noise level International standard: 140 dB relative to 20 mPa,40,Safety (continued),-Contrast agents -Gd-DTPA is significantly safer than iodine based contrast agents used in X-ray procedures (x 10) -Severe reactions do occur -Pregnancy: Gd-DTPA may
34、have poor clearance from amniotic fluid,41,Safety (continued),-Psychological Distress (before, during and after MRI) Claustrophobia Anxiety Depression Panic Disorders -RF coils need to be near the body for high signal to noise -Usually transient (2 cases of claustrophobia persisted requiring psychia
35、tric treatment),42,Safety (continued),-Psychological Distress -Countermeasures Briefing Friend or relative in scan room Physical or verbal contact with patient Prone Mirrors, prism glasses, video goggles Blindfold Relaxation strategies Psychological desensitization techniques,43,Cryogen Safety,Super
36、conducting MRI magnets use cryogens (1970 liters of liquid helium) Magnets can QUENCH resulting in conversion of the liquid helium to gaseous helium 1970 liters of liquid helium becomes 53,000 cubic feet of gaseous helium in about 1 minute,44,Cryogen Safety,Magnets are designed to dump the gaseous h
37、elium directly outdoors through a quench pipe If the quench pipe fails, gaseous helium could be dumped into the scan room, displacing the oxygen MCGHI MRI scan rooms have oxygen monitors that automatically activate emergency exhaust systems in the event the oxygen level drops below 18% (An Alarm Sou
38、nds),45,Cryogen Safety,There has been one fatality by asphyxiation related to a liquid nitrogen leak in MRI,46,Safety (continued),-Experiences with higher magnetic fields (4 T) A few 7T systems Several 4 T systems Numerous 3T systems Bioeffects: magnetophosphenes vertigo nausea headaches metallic taste,47,Pregnancy,MRI is not considered to be hazardous to the fetus Limited studies have examined teratogenic effects Cells that are reproducing rapidly, as in the first trimest
温馨提示
- 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
- 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
- 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
- 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
- 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
- 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
- 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。
最新文档
- 网商岗前风险识别考核试卷含答案
- 木材削片工安全技能知识考核试卷含答案
- 甲烷合成气净化工诚信水平考核试卷含答案
- 2026银行it岗半结构化面试题及答案
- 2026银行考官面试题及答案
- 2026银行行长面试题及答案
- 2026印刷行业面试题目及答案
- 长春部分学校2026-2027学年高三上学期第一次摸底考试语文试题及参考答案
- (2026年)超声检查与解读报告基础知识
- 2026慈溪联考语文面试题目及答案
- 转轮除湿机选型设计计算书
- ISO9001-2026质量管理体系“气候变化”要求专项实施操作指导材料(雷泽佳编制-2026A1)
- 2026年团员入团知识测试题(各地真题)
- 船舶新能源技术
- 2025国家基层糖尿病防治管理指南解读课件
- 林木居间协议合同范本
- 现浇梁蒸养棚施工方案
- 地下室顶板加固施工方案范本
- 门禁工程施工方案
- 厦门安辛安全生产培训课件
- 未成年禁止骑摩托安全协议书5篇
评论
0/150
提交评论